Standard Calibration
RAPT Pill Hydrometer & Thermometer - Standard Calibration Guide
Standard calibration mode is recommended for most users.
Standard calibration is the quickest and simplest method of calibration.
Enter Calibration mode
Plug your Rapt Pill into a charger or computer with a USB-C cable (not included) or place on the wireless charger (if the wireless charging kit is installed)
On your smart phone or tablet, open your WiFi settings and connect to the RAPT Pill XXXX access point using the password "kegland1". You may not need to enter the password if you have previously connected your device to the Pill WiFi
Once the RAPT Pill Access Portal is open on your phone or tablet, navigate to the Calibration tab
Perform Standard Calibration
Fill a container of sufficient volume to hold the RAPT Pill in suspension with a liquid of known gravity. It is simplest and easiest to use water which has a gravity of 1.000
Place the Pill gently into the liquid. Ensure that it can float freely and is not touching the bottom of the vessel, as this will affect the angle
Select Standard Calibration in the Pill Captive Portal
Enter the known gravity of the liquid - typically 1.00 for water or sanitiser at the correct dilution
Wait for the angle reading to stabilise. The angle will never be completely steady, as it is extremely sensitive
Hit Calibrate and wait for Calibration to be completed
Your Pill is now calibrated and ready to use!
Last updated