Calibration
RAPT Pill Hydrometer & Thermometer - Calibration Guide
Last updated
RAPT Pill Hydrometer & Thermometer - Calibration Guide
Last updated
Calibration before first use is strongly encouraged for best results.
It is recommended to recalibrate after the Pill housing has been opened (to recharge via USB-C, for example), as tiny changes when opening/closing the housing can affect calibration.
IMPORTANT: The RAPT Pill is designed to disconnect from the RAPT Portal once the USB-C cable is unplugged to preserve battery life. Opening the Calibration or Diagnostics screen before disconnecting will keep the Pill Captive Portal active to allow calibration of the Pill. You will be prompted after 10 minutes to “Click Here To Maintain Connection” if the calibration process takes longer than 10 minutes.
Once you have entered a Calibration mode, please assemble your Pill according to the directions so that it can be removed from the USB and calibrated according to the instructions below
Plug your Rapt Pill into a charger or computer with a USB-C cable (not included) or place on the wireless charger (if the wireless charging kit is installed)
On your smart phone or tablet, open your WiFi settings and connect to the RAPT Pill XXXX access point using the password "kegland1". You may not need to enter the password if you have previously connected your device to the Pill WiFi
Once the RAPT Pill Access Portal is open on your phone or tablet, navigate to the Calibration tab