LogoLogo
Registration GuideTroubleshoot RegistrationProfiles Guide
  • RAPT Portal - Introduction
  • Getting Started
  • Dashboard
  • Equipment
    • Devices
    • RAPT BrewZilla
      • Get Started!
      • Setup & Assembly
      • Manual Operation with the Digital Controller
      • Remote Operation with the RAPT Portal
      • Brewing with RAPT Gen 4 BrewZilla
      • Using the RAPT Bluetooth Thermometer
      • Profiles
      • Menus
      • Troubleshoot BrewZilla
      • Convert Single Phase to Three Phase (100L Gen 4 BrewZilla only)
      • RAPT Device Boot Loop Troubleshooting & Recovery Mode
    • RAPT Can Filler
      • Getting Started
      • Setup
      • Operation
      • Cleaning
      • Antenna Upgrade
      • Installing KL38171 Instant Read Water Resistant Probe
      • Installing KB22681 Slimline Fill Tube Kit
      • Troubleshooting Can Filler
      • RAPT Device Boot Loop Troubleshooting & Recovery Mode
    • RAPT Fermentation Chamber
      • Get Started!
      • Setup
      • Operation
      • Profiles
      • Menus
      • Bluetooth
      • Screen Change For RAPT Fermentation Chamber
      • Troubleshooting
      • RAPT Device Boot Loop Troubleshooting & Recovery Mode
    • RAPT Pill Hydrometer
      • Get Started!
      • RAPT Pill Captive Portal
      • Calibration
        • Standard Calibration
        • Advanced Calibration
        • Custom Calibration
      • Operation
      • Bluetooth Operation
      • Pill Wireless Charger Upgrade - Installation
      • Troubleshooting
    • RAPT Temperature Controller
      • Get Started!
      • Setup
      • Operation
      • Profiles
      • Menus
      • Bluetooth
      • Troubleshooting
      • RAPT Device Boot Loop Troubleshooting & Recovery Mode
    • RAPT Bluetooth Thermometer
      • Assembly & Setup
      • Bluetooth Operation
      • Settings
  • Profiles
    • Starting or Stopping a Profile
    • Creating and Editing Profiles
      • BrewZilla Profiles
        • Advanced BrewZilla Profiles
      • General Temperature Control Profiles
      • Pill Hydrometer Profiles
      • Fermentation Profiles
    • My Profiles
    • Search Profiles
    • Device Profile Sessions
      • Profile Session Time Travel (Skip Steps)
  • Tools
    • ABV Calculator
    • Colour Converter
    • Wort Dilution Calculator
    • Hydrometer Temperature Adjustment Calculator
  • My Account
  • Troubleshooting
Powered by GitBook
On this page
  • Enter Calibration mode
  • Perform Advanced Calibration
  1. Equipment
  2. RAPT Pill Hydrometer
  3. Calibration

Advanced Calibration

RAPT Pill Hydrometer & Thermometer - Advanced Calibratin Guide

Advanced calibration mode is suited for higher gravity ferments.

Advanced calibration mode can give greater accuracy across a wider range. Advanced calibration is also quite quick and easy, and can provide a more accurate result than Standard Calibration.

It is the same procedure as standard calibration, but in addition to water at 1.000, you need a solution of a known gravity prepared. We recommend using a gravity in the range that your fermentation will start – 1.040-1.060 for typical beers, for example.

Enter Calibration mode

  • Plug your Rapt Pill into a charger or computer with a USB-C cable (not included) or place on the wireless charger (if the wireless charging kit is installed)

  • On your smart phone or tablet, open your WiFi settings and connect to the RAPT Pill XXXX access point using the password "kegland1". You may not need to enter the password if you have previously connected your device to the Pill WiFi

  • Once the RAPT Pill Access Portal is open on your phone or tablet, navigate to the Calibration tab

Perform Advanced Calibration

  • Fill a container of sufficient volume to hold the RAPT Pill in suspension with a liquid of known gravity. It is simplest and easiest to use water or sanitiser which has a gravity of 1.000

  • Fill another container capable of holding the Pill in suspension with a liquid of known gravity in the range representative of the upper limit of your starting gravity

  • Place the Pill gently into the weaker solution. Ensure that it can float freely and is not touching the bottom of the vessel, as this will affect the angle

  • Select Advanced Calibration in the Pill Captive Portal

  • Enter the known gravity of the liquid - typically 1.00 for water or sanitiser at the correct dilution

  • Wait for the angle reading to stabilise. The angle will never be completely steady, as it is extremely sensitive

  • Hit Calibrate and wait for the first calibration step to be completed

  • Place the Pill in the stronger solution. Enter the gravity as determined by a calibrated hydrometer. Once the angle reading has stabilised press Calibrate. Wait until the screen indicates that calibration is complete

Your Pill is now calibrated and ready to use!

PreviousStandard CalibrationNextCustom Calibration