Calibration
RAPT Can & Bottle Filler | Calibration Guide
Please note the calibration process has changed since a firmware update on the 30/04/2025.
Firmware release date can be identified by the first 8 digits of the firmware:
Version: 20250430_155724_ed6e01f was released on 30/04/2025, for example
Ensure you are running firmware from this date or newer before calibrating for the first time.
Prepare the Machine
Mark the Purge Tube: Locate the stainless steel purge dip tube (on the right side of the unit). Measure 100mm up from the bottom of the tube and mark this clearly with a wipeable marker.
Connect Gas: Attach your gas line to the red carbonation cap. This is required to clear the line during the calibration process.
Set the 2nd Calibration Point (Reference Height)
Prepare Reference Water: Place a container (a glass is ideal) under the fill & purge tubes. Fill it with water until the water level exactly hits the 100mm mark you made in Step 1.
Check Value: Ensure the measurement value on the screen matches the height of your water (e.g., 100mm). If you used a different height, update the value on the screen now.
Capture Reading: Wait for the pressure sensor value to stabilise (fluctuating by only a few points). Select
. The unit now knows the pressure-to-depth ratio.
Run a Test Fill With Water
Set Your Target: Place your actual can or bottle in the filling position. Decide where you want the liquid to stop and mark that spot on the purge tube.
Measure & Program: Measure from the bottom of the tube up to this new mark. Enter this value (in mm) into your Fill Preset.
Test: Connect your water source to the yellow carbonation cap (max 15psi). Select your preset, place your container, and press Run.
Last updated