# Calibration

KL20596 RAPT Pill Hydrometer & Thermometer | Calibration

**Calibration before first use is strongly encouraged for best results.**

**It is recommended to recalibrate after the Pill housing has been opened (to recharge via USB-C, for example), as tiny changes when opening/closing the housing can affect calibration.**

Calibration is performed through the **RAPT Brew App**

**RAPT Pill Hydrometers must be running an up to date firmware in order to be Calibrated directly through the app.**

If there is no Calibration option on the Pill Device screen, the Pill may need to be updated. To update the Pill firmware, simply place it on a charger if wireless, or connect via USB-C. RAPT Pills will automatically check and update firmware when connected to a WIFi network and charging.

To see the **Calibrate** option, you may need to swipe the **Device** page to the right as indicated:

<figure><img src="https://570278281-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FUqN4sW0x2s4yh6iiT8NN%2Fuploads%2FQQjZyUOm1I9vNT8dWM7D%2FCalibrate.jpg?alt=media&amp;token=d1ed73ac-e5ef-498b-8dd9-1ae34173f9c4" alt=""><figcaption></figcaption></figure>

The Pill must be connected to the RAPT Brew App to Calibrate. The Pill needs to be in **Smart Mode** in order to connect. Smart Mode is the default mode.

### Calibration

{% embed url="<https://www.youtube.com/watch?t=564s&v=jv7Ditichko>" %}

#### Connect Pill

First, **Connect to Pill**. This will take up to 30 seconds in **Smart Mode**

<figure><img src="https://570278281-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FUqN4sW0x2s4yh6iiT8NN%2Fuploads%2FyEXKEMTm8kiDvZU0GNRg%2FConnect%20to%20pill.jpg?alt=media&amp;token=ef74330b-4782-453d-b1cf-222673b94729" alt=""><figcaption></figcaption></figure>

Once Connected, the button will change to **Refresh Reading**. The Green LED on the RAPT Pill will display solid green, indicating a direct Bluetooth connection.

There are three Calibration options - **Water Calibration, Enter Known Gravity, Forget Calibration**

#### Water Calibration

Water Calibration allows you to quickly and easily calibrate the Pill in water or sanitiser. It will work with any liquid with a Specific Gravity of 1.000.

[Connect to the Pill](#connect-pill)

Once connected, place the RAPT Pill gently in a jug of water or sanitiser (or any vessel large enough to allow the Pill to float freely) and select Water Calibration.

Select **Confirm** to calibrate. The display will change to show **Calibration** is underway.

<figure><img src="https://570278281-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FUqN4sW0x2s4yh6iiT8NN%2Fuploads%2Fj8Fv4weQ37EuO0z1qrXO%2Fimage.png?alt=media&amp;token=e5f09bc8-6521-48c8-968e-fe912641617a" alt=""><figcaption></figcaption></figure>

**Calibration Successful** will be shown once **Calibration** is complete, and the displayed gravity will change to reflect the new calibration.

#### Enter Known Gravity

Enter Known Gravity allows you to Calibrate the RAPT Pill in a solution of known gravity (such as wort which has been measured with a calibrated hydrometer). For accurate calibration, this solution should have an SG of 1.020 or higher.

[Connect to the Pill](#connect-pill)

Once connected, place the RAPT Pill gently in a jug of solution or wort of a known specific gravity (or any vessel large enough to allow the Pill to float freely) and select Enter Known Gravity.

Enter the Known Gravity in the displayed field.

<figure><img src="https://570278281-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FUqN4sW0x2s4yh6iiT8NN%2Fuploads%2FK1sd8PJgD1tYM78WLFUC%2Fimage.png?alt=media&amp;token=4c3eefc8-773b-4626-8796-bf88516ebe5d" alt=""><figcaption></figcaption></figure>

Select **Confirm** to calibrate. The display will change to show **Calibration** is underway.

**Calibration Successful** will be shown once **Calibration** is complete, and the displayed gravity will change to reflect the new calibration.

**NOTE**: Only one Known Gravity calibration point is stored.

#### Forget Calibration

Use this option to forget the current calibration if required. This removes both calibration points

### Calibration Notes

Calibration is not time sensitive, and the Calibration settings are persistent until overwritten. This means you can easily get very accurate calibration in real world settings.

A great way to calibrate the RAPT Pill is when pitching yeast in a fermenter. Place the RAPT Pill in the fermenter or other vessel that has sufficient santiser to allow the Pill to float freely, and perform Water Calibration. This sets the first Calibration point, and also sanitises the Pill prior to use.

Next, once the wort is in the fermenter, drop the Pill into the Fermenter and select Enter Known Gravity. Enter the known gravity of the wort and Calibrate. this is the second calibration point.

Note: Water Calibration is generally quite accurate. Using two calibration points will give more accurate results, especially in higher gravity (>1.064) worts.