# Calibration

## <img src="/files/SgrkYdQlWtKd0MmS25WV" alt="" data-size="line"> Mandatory Firmware Check

{% hint style="info" %}
Please Note: The calibration process was significantly updated on 30/01/2025. To ensure your device operates correctly and follows the steps in this manual, you must be running this firmware version or newer.
{% endhint %}

{% stepper %}
{% step %}

### Prepare the Machine

**Mark the Purge Tube:** Locate the stainless steel purge dip tube (on the right side of the unit). Measure 100mm up from the bottom of the tube and mark this clearly with a wipeable marker.

**Connect Gas:** Attach your gas line to the red carbonation cap. This is required to clear the line during the calibration process.
{% endstep %}

{% step %}

### Enter Calibration Mode

To enter the calibration mode: **Select** <img src="https://docs.rapt.io/~gitbook/image?url=https%3A%2F%2F570278281-files.gitbook.io%2F%7E%2Ffiles%2Fv0%2Fb%2Fgitbook-x-prod.appspot.com%2Fo%2Fspaces%252FUqN4sW0x2s4yh6iiT8NN%252Fuploads%252FJVin8Ct47ODMRC9a2qab%252FENTER%2520RAPT.png%3Falt%3Dmedia%26token%3D077fda61-812a-4aaa-8693-2e8d66c23e2e&#x26;width=40&#x26;dpr=4&#x26;quality=100&#x26;sign=731735a3&#x26;sv=1" alt="" data-size="line"> then navigate through the menu to **Calibrate** then press **Fill**.
{% endstep %}

{% step %}

### Set the 1st Calibration Point (Ambient/Empty)

**Clear the Line:** Remove the purge tube from container or liquid. Press **Select** <img src="https://docs.rapt.io/~gitbook/image?url=https%3A%2F%2F570278281-files.gitbook.io%2F%7E%2Ffiles%2Fv0%2Fb%2Fgitbook-x-prod.appspot.com%2Fo%2Fspaces%252FUqN4sW0x2s4yh6iiT8NN%252Fuploads%252FJVin8Ct47ODMRC9a2qab%252FENTER%2520RAPT.png%3Falt%3Dmedia%26token%3D077fda61-812a-4aaa-8693-2e8d66c23e2e&#x26;width=40&#x26;dpr=4&#x26;quality=100&#x26;sign=731735a3&#x26;sv=1" alt="" data-size="line">.&#x20;

**Capture Reading:** Press **Select** <img src="https://docs.rapt.io/~gitbook/image?url=https%3A%2F%2F570278281-files.gitbook.io%2F%7E%2Ffiles%2Fv0%2Fb%2Fgitbook-x-prod.appspot.com%2Fo%2Fspaces%252FUqN4sW0x2s4yh6iiT8NN%252Fuploads%252FJVin8Ct47ODMRC9a2qab%252FENTER%2520RAPT.png%3Falt%3Dmedia%26token%3D077fda61-812a-4aaa-8693-2e8d66c23e2e&#x26;width=40&#x26;dpr=4&#x26;quality=100&#x26;sign=731735a3&#x26;sv=1" alt="" data-size="line">. The unit will purge the line and display `Purging > Waiting for Pressure to Stabilize`. This sets the baseline "zero" value for ambient pressure.
{% endstep %}

{% step %}

### Set the 2nd Calibration Point (Reference Height)

**Prepare Reference Water:** Place a container (a glass is ideal) under the fill & purge tubes. Fill it with water until the water level exactly hits the 100mm mark you made in Step 1.

**Check Value:** Ensure the measurement value on the screen matches the height of your water (e.g., 100mm). If you used a different height, update the value on the screen now.

**Capture Reading:** Wait for the pressure sensor value to stabilise (fluctuating by only a few points). **Select** <img src="https://docs.rapt.io/~gitbook/image?url=https%3A%2F%2F570278281-files.gitbook.io%2F%7E%2Ffiles%2Fv0%2Fb%2Fgitbook-x-prod.appspot.com%2Fo%2Fspaces%252FUqN4sW0x2s4yh6iiT8NN%252Fuploads%252FJVin8Ct47ODMRC9a2qab%252FENTER%2520RAPT.png%3Falt%3Dmedia%26token%3D077fda61-812a-4aaa-8693-2e8d66c23e2e&#x26;width=40&#x26;dpr=4&#x26;quality=100&#x26;sign=731735a3&#x26;sv=1" alt="" data-size="line">. The unit now knows the pressure-to-depth ratio.
{% endstep %}

{% step %}

### Run a Test Fill With Water

**Set Your Target:** Place your actual can or bottle in the filling position. Decide where you want the liquid to stop and mark that spot on the purge tube.

**Measure & Program:** Measure from the bottom of the tube up to this new mark. Enter this value (in mm) into your Fill Preset.

**Test:** Connect your water source to the yellow carbonation cap (max 15psi). Select your preset, place your container, and press Run.
{% endstep %}
{% endstepper %}

<table data-view="cards"><thead><tr><th data-hidden data-card-cover data-type="image">Cover image</th><th data-hidden data-card-target data-type="content-ref"></th></tr></thead><tbody><tr><td><a href="/files/zd0iIewJyThD03ujbysb">/files/zd0iIewJyThD03ujbysb</a></td><td><a href="https://youtu.be/BZ8N-igzbTM?si=pAt4P460Uk-ikJFR&#x26;t=305">https://youtu.be/BZ8N-igzbTM?si=pAt4P460Uk-ikJFR&#x26;t=305</a></td></tr></tbody></table>


---

# Agent Instructions: Querying This Documentation

If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter:

```
GET https://docs.rapt.io/equipment/rapt-can-filler-or-user-guide/calibration.md?ask=<question>
```

The question should be specific, self-contained, and written in natural language.
The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.